A Journal of Analytical Chemistry and Chemical Analysis
Yazarlar: Merve Yence, Leyla Karadurmus, Goksu Ozcelikay, Nurgul K. Bakirhan, Sibel A. Ozkan
Konular:-
DOI:10.3390/analytica2040017
Anahtar Kelimeler:Ivabradine hydrochloride,Nanosensor,Glassy carbon electrode,Mesoporous carbon solution,COOH functionalized SWCNT
Özet: In this study, the electrochemical determination of Ivabradine hydrochloride (IH) was studied in detail using a glassy carbon electrode (GCE) modified with mesoporous carbon solution (MCS) and carboxylated group linked single-walled carbon nanotube (SWCNT-COOH). The developed nanosensor showed a significant effect by remarkably increasing the IH signal compared with the bare GCE. Cyclic (CV) and differential pulse voltammetric (DPV) methods were applied to perform electrochemical analysis of IH in pH 3.0 BRB solutions. The calibration plot for IH with a detection limit of 1.47 × 10 −7 M was obtained using the DPV technique in the range of 1–10 µM under optimum experimental conditions. The proposed method has been validated and applied for the detection of the IH tablet. The produced nanosensor was also performed for the determination of IH in serum and urine. Excellent recoveries of 98.4%, 98.0%, and 100.2% were achieved for tablet, serum, and urine analysis, respectively.