European Endodontic Journal
Yazarlar: Emmanuel João Nogueira Leal Silva, Justine Monnerat Tinoco, Eduardo Muniz Barretto Tinoco, Victor Talarico Leal Vieira, Luciana Moura Sassone, Hélio Pereira Lope
Konular:-
DOI:10.5152/eej.2016.16018
Anahtar Kelimeler: Bending resistance,Cyclic fatigue,Instrumentation and rotary,Reciprocating movement,NiTi alloy,WaveOne Gold
Özet: The present study aimed to evaluate the bending resistance and cyclic fatigue fracture resistance of a new single-file reciprocating instrument called WaveOne Gold. Reciproc and WaveOne instruments were used as references for comparison. Methods: Sixty 25-mm NiTi instruments (Reciproc R25, WaveOne Primary and WaveOne Gold Primary) were tested. Flexibility was determined by applying 45° bending tests using a universal testing machine (n=10). A custom-made device was used during cyclic fatigue test (n=10), comprising a stainless steel artificial ca- nal measuring 1.4 mm in diameter, 19 mm in total length with an 86° angle and 6 mm radius of curvature. Possible deformations at the helical shaft and mode of fracture were evaluated using scanning electron mi- croscopy analysis. Statistical analysis was performed using one-way analysis of variance. Post hoc pair-wise comparisons were performed using Tukey’s test for multiple comparisons (P<0.05). Results: WaveOne presented significantly higher bending resistance than the other tested systems (P<0.05), whereas Reciproc presented the lowest bending resistance (P<0.05). Reciproc revealed a significantly longer cyclic fatigue fracture resistance than the other systems (P<0.05). Conclusions: Although WaveOne Gold presented higher flexibility than WaveOne, no differences in the resis- tance to fatigue were observed between both systems. The Reciproc files were more flexible and resistant to fatigue for the angle of curvature of 86° and 6 mm radius than WaveOne and WaveOne Gold files.